What Is the Calibration of a Temperature and Humidity Test Chamber? A Comprehensive Guide to Measuring and Evaluating the Performance of Environmental Test Chambers

What Is the Calibration of a Temperature and Humidity Test Chamber? A Comprehensive Guide to Measuring and Evaluating the Performance of Environmental Test Chambers

Environmental test chambers are used to create and control conditions such as temperature and humidity within a specified space. These devices are used in research testing, quality control, product durability evaluation, and environmental condition simulation across various industries.

However, one fundamental question arises:

How can we be sure that the temperature and humidity displayed by a chamber correspond to the actual conditions inside the chamber?

To answer this question, the condition of the measurement system and the performance of the chamber’s working space must be examined.

In this article, we examine the concept of calibration, its difference from adjustment and performance verification, methods for evaluating temperature and humidity, measurement error, uniformity and stability, reference equipment, uncertainty, traceability, and the standards related to environmental test chambers.

What Is Chamber Calibration?

First, an important point must be clarified: in industry, the term “chamber calibration” may be used to refer to several different activities.

From a metrological perspective, Calibration is a measurement process in which the relationship between values provided by a reference and the values indicated or generated by the system under examination is determined.

In simpler terms, calibration determines:

What value does the device indicate, and what value does the reference measure?

For example, suppose the controller of a chamber indicates a temperature of 80.0°C. Under the same conditions, a reference measurement system records a value of 79.6°C.

In this case, the difference between the two values is:

0.4°C.

However, this difference alone does not indicate whether the chamber is acceptable or not. To determine whether the result is acceptable, an acceptance criterion must be specified.

This criterion may be derived from the device’s technical specifications, the test method, customer requirements, the laboratory’s internal procedure, or the standard relevant to the application.

Therefore, calibration should not simply be defined as “adjusting the device” or “correcting the display reading.”

Is Chamber Calibration the Same as Sensor Calibration?

No; and this is one of the most important points in evaluating test chambers.

Suppose the PT100 sensor inside a chamber has been calibrated separately and has a valid calibration certificate. This indicates that the measurement performance of that sensor has been examined under specified conditions, but it does not, by itself, prove that the entire working space of the chamber has a uniform temperature.

In actual chamber performance, many more factors are involved, including:

  • Temperature sensor
  • Sensor installation location
  • Controller
  • Heating system
  • Cooling system
  • Fan and air circulation system
  • Airflow path
  • Internal space design
  • Insulation
  • Door sealing
  • Wall temperature
  • Load inside the chamber
  • Ambient conditions

For this reason, a distinction must be made between sensor calibration and the evaluation of the chamber’s working space performance.

A sensor may be highly accurate, but the airflow design may cause the temperatures at different corners of the chamber to differ from one another.

Why Are Chamber Calibration and Performance Evaluation Important?

The purpose of using a chamber is to create specific and controllable environmental conditions for the test specimen.

If the actual test conditions differ from the expected conditions, the test result may also be affected.

For example, suppose an electronic component must be tested at 85°C for a specified period. If the actual temperature of the specimen is lower or higher than the target value due to temperature differences within the working space, the actual test conditions will not be identical to the defined conditions.

The importance of evaluating chamber performance becomes greater in the following applications:

  • Quality control testing
  • Durability and lifetime testing
  • Environmental testing of electronic products
  • Automotive component testing
  • Aerospace equipment testing
  • Medical equipment testing
  • Pharmaceutical testing
  • Materials and polymer testing
  • Research testing
  • Processes subject to quality and documentation requirements

In such applications, it is not sufficient for the chamber merely to “reach the set temperature”; it must also be determined to what extent the conditions created within the chamber’s working space comply with the test requirements.

What Is the Difference Between Calibration, Verification, and Adjustment?

The three terms Calibration, Verification, and Adjustment are sometimes used interchangeably in industry, although they are technically different.

Calibration

In calibration, the relationship between the reference value and the value indicated or obtained from the measurement system is determined.

The result may include measured values, error, correction, and, where necessary, information related to uncertainty.

Verification

Verification means providing objective evidence that an equipment item or process meets specified requirements.

For example, if it has been specified that the temperature error of a chamber must not exceed ±1°C, the measurement result is compared with this criterion.

Therefore:

Calibration determines the measurement status; Verification examines whether this status meets the specified requirements.

The definition of Verification in the International Vocabulary of Metrology (VIM) also emphasizes this concept of “providing objective evidence that specified requirements have been fulfilled.”

Adjustment

Adjustment refers to the process of correcting or adjusting a measurement or control system so that its performance is brought closer to the desired value.

What Is the Calibration of a Temperature and Humidity Test Chamber? A Comprehensive Guide to Measuring and Evaluating the Performance of Environmental Test Chambers

What Is the Calibration of a Temperature and Humidity Test Chamber? A Comprehensive Guide to Measuring and Evaluating the Performance of Environmental Test Chambers

What Parameters Are Evaluated in Test Chambers?

The parameters examined depend on the type of chamber and its application.

In a temperature test chamber, the most important parameter is usually temperature.

In a temperature and humidity chamber, relative humidity is also important in addition to temperature.

Other environmental test equipment may require additional parameters, such as:

  • Temperature
  • Relative humidity
  • Pressure
  • Irradiance
  • UV light intensity
  • Air velocity or airflow conditions
  • Other parameters defined in the test method

Therefore, a single method cannot be defined for the calibration and performance evaluation of all types of chambers.

Four Important Concepts in Chamber Performance Evaluation

To evaluate the performance of a chamber accurately, several concepts must be distinguished from one another:

  • Indication Error
  • Stability
  • Uniformity
  • Measurement Uncertainty

Confusing these concepts is one of the common reasons for misinterpreting chamber measurement results.

What Is Temperature Indication Error?

Suppose the chamber controller indicates the following value:

80.0°C

Under the same conditions, the reference measurement system records the following value:

79.6°C

The difference between the two values is:

0.4°C

This difference can be examined as one of the results of calibrating the indication/measurement system.

However, the important point is that the chamber’s performance should not be judged solely based on the value of 0.4°C.

For example, if the process acceptance criterion is ±1°C, this value may be acceptable. If the acceptance criterion is ±0.2°C, it may not be acceptable.

Therefore:

Measurement error is not the same as equipment nonconformity.

The acceptance criterion must first be specified.

What Is Temperature Stability?

Suppose the temperature at a specific point in the chamber is recorded over a period of time as follows:

80.0 → 80.2 → 79.9 → 80.1 → 80.0°C

Here, the temperature has changed over time.

Examining these changes is part of temperature stability evaluation.

Therefore, the main question in Stability is:

How much does the temperature at a specific point change over time?

Stability is related to the temporal behavior of temperature.

What Is Temperature Uniformity?

Uniformity is different from Stability.

In Uniformity, instead of examining the changes at one point over time, we compare the temperatures at different points within the working space.

For example, suppose the temperatures at different points in the chamber at a specific time are as follows:

Upper left: 80.4°C

Upper right: 79.7°C

Center: 80.0°C

Lower left: 79.3°C

Lower right: 79.8°C

In this example, the chamber display may indicate 80°C, but the entire working space is not exactly at 80°C.

Therefore:

Stability: Changes over time

Uniformity: Differences between different spatial locations

These two parameters must be evaluated separately.

Why Is Temperature Uniformity Important in a Chamber?

Suppose we place two identical specimens in two different locations within a chamber.

If one specimen is located at a point with a temperature of 79.3°C and the other at a point with a temperature of 80.4°C, the thermal conditions of the two specimens are not exactly the same.

This issue can be important in sensitive tests.

For this reason, a suitable chamber must not only be capable of reaching the target temperature; it must also be able to create the required conditions throughout the defined working space with appropriate uniformity and stability.

Is the Temperature Displayed by the Chamber the Same as the Actual Temperature of the Specimen?

Not necessarily.

This is one of the important points to consider when using a chamber.

The controller sensor is usually located at a specific position inside the device. The controller regulates the heating or cooling system based on that measurement.

However, the specimen may be located at another position.

As a result, the following may occur:

Control sensor temperature = 80°C

But:

Specimen location temperature = 79.5°C

Of course, the temperature difference depends on the chamber design, airflow, operating conditions, sensor location, temperature and load inside the chamber, and other factors.

How Is Chamber Calibration Performed?

The exact procedure depends on the objective, chamber type, measured parameter, and the standard or method being used.

However, a systematic process usually includes the following steps.

  1. Initial Chamber Inspection

Before starting the measurements, the general condition of the chamber must be examined.

For example:

  • Controller operation
  • Sensor operation
  • Heating system
  • Cooling system
  • Fan
  • Humidification system
  • Dehumidification system
  • Door sealing
  • Condition of the internal space
  • Ambient conditions

If the chamber has a technical problem, calibration cannot replace repair.

  1. Determining the Measurement Range and Points

Before carrying out the work, it must be specified which conditions are to be examined.

For example, the objective may be to evaluate chamber performance at several temperature points.

The number and locations of measurement points must be determined based on the method and the evaluation objective.

Therefore, statements such as:

“Every chamber must be calibrated at exactly 9 points” or “Every chamber always requires 15 sensors” are not scientifically correct as general rules.

The number of points depends on the dimensions of the working space, the test method, the standard being used, and the measurement objective.

  1. Selecting Reference Equipment

For temperature measurement, suitable reference systems may be used, such as:

  • Reference thermometers
  • Reference RTD sensors
  • PT100
  • Suitable thermocouples
  • Data Logger
  • Multichannel recording systems

Reference humidity measurement equipment is used for humidity measurements as well.

However, an important point must be considered:

The reference equipment itself must have a clearly established metrological status.

The fact that a Data Logger or PT100 is accurate is not, by itself, sufficient to produce a valid calibration result.

The operating range, equipment specifications, permissible error, uncertainty, and traceability must be compatible with the intended application.

  1. Sensor Placement

Sensor placement has a direct effect on the result.

If a sensor:

  • Is placed too close to a wall,
  • Is located directly in the path of the outlet air,
  • Is placed near a heat source,
  • Is positioned at an unsuitable location within the working space,

the measured value may not represent the actual conditions of the space being examined.

Therefore, sensor placement must be determined according to the measurement method and should not be selected merely for convenience.

  1. Allowing the Chamber to Reach the Appropriate Conditions

After setting the temperature or humidity, the data cannot be immediately recorded as the final result.

The chamber must reach the required conditions, and its behavior must be examined over the defined time interval.

This step is very important because the heating, cooling, humidity, and airflow systems may require time to reach a stable state.

  1. Recording Data

During the process, the data from the reference equipment and the value indicated by the chamber are recorded.

Using a multichannel Data Logger has the advantage of allowing the behavior of several points to be recorded simultaneously and analyzed later.

  1. Calculating and Analyzing the Results

After collecting the data, the results are examined.

Items such as:

  • Difference between the reference value and the display
  • Temporal variations
  • Differences between different points
  • Measurement uncertainty

may be considered in the final analysis.

At this stage, it is important that the result is not presented merely as several raw numbers; the meaning of each number must also be clear.

Is Chamber Calibration Always Performed in the Empty Condition?

No; this depends on the type of evaluation.

In some chamber performance verification methods, conditions without a specimen or load are used.

On the other hand, for some tests, evaluating conditions under load is important.

For example, IEC 60068-3-7:2020 addresses measurements in temperature chambers for cold and dry heat tests with load.

Therefore, a general rule such as “calibration is always performed with a load” or “calibration is always performed without a load” cannot be provided.

The objective of the measurement must first be specified.

How Many Points Are Required for Chamber Calibration?

This is a very common question, but the answer is not a fixed number.

The number of points depends on factors such as:

  • Dimensions of the working space
  • Chamber geometry
  • Chamber type
  • Operating range
  • Measured parameter
  • Test objective
  • Standard being used
  • Measurement method
  • Customer requirements

Consequently, the number of points must be determined based on the selected method.

This is also important for the chamber manufacturer, because the device must be designed in such a way that the required operating points and airflow conditions allow its performance to be evaluated appropriately.

How Does Humidity Chamber Calibration Differ from Temperature Calibration?

Relative humidity is a parameter that depends on temperature.

For this reason, evaluating the humidity of a chamber does not simply mean placing an RH sensor inside the chamber and reading the display.

Temperature and humidity must be examined together and in accordance with the measurement method.

In temperature and humidity chambers, factors such as the following may affect the final performance:

  • Air temperature
  • Relative humidity
  • Humidity stability
  • Humidity uniformity
  • Airflow conditions
  • Condition of the humidification system
  • Dehumidification system
  • Condensation on surfaces

For this reason, from a measurement perspective, evaluating a temperature and humidity chamber is generally more complex than evaluating a temperature-only chamber.

What Is Measurement Traceability?

When discussing calibration, the term Metrological Traceability is very important.

According to the VIM, metrological traceability is a property of a measurement result whereby the result can be related to a reference through a documented and unbroken chain of calibrations, each of which contributes to the measurement uncertainty.

In simple terms, if we use a reference thermometer to examine a chamber, it must be clear how and against which reference that instrument has been calibrated.

A simple chain can be represented as follows:

Chamber measurement result ← Reference equipment ← Reference standard ← Higher-level reference

The objective is to ensure that the measurement result is not merely a number without supporting evidence.

What Is Measurement Uncertainty?

In no real measurement system can we say that the result is completely free of uncertainty.

Suppose a reference instrument indicates a value of 80.0°C.

This does not mean that the actual value is exactly and unambiguously 80.000°C.

Various factors may contribute to the uncertainty of the result, including:

  • Specifications of the reference equipment
  • Calibration of the reference equipment
  • Display resolution
  • Repeatability
  • Stability of the conditions
  • Measurement method
  • Sensor location
  • Ambient conditions
  • Temperature gradients
  • Thermal and radiative effects
  • Other components of the measurement model

Therefore, in a professional process, reporting only the error value is not sufficient; depending on the type of work, uncertainty must also be taken into account.

Does an Error of 0.5 Degrees Mean That the Chamber Is Rejected?

No.

Suppose the measurement result indicates:

Error = +0.5°C

This number alone cannot determine the status of the chamber.

We must know:

What is the acceptance criterion?

If the acceptance criterion is ±1°C, this value may be acceptable.

If the acceptance criterion is ±0.2°C, it may not be acceptable.

Even in a more professional interpretation, the method of deciding conformity and the role of measurement uncertainty must also be specified according to the relevant method and decision rules.

Therefore, the statement “The device has an error of 0.5 degrees, so it is not accurate” is technically incomplete.

What Information Does a Chamber Calibration Certificate Include?

A suitable calibration report or measurement report should provide sufficient information for the reader to understand the result.

Depending on the type of process, this information may include:

  • Equipment name
  • Model
  • Serial number
  • Range examined
  • Measurement date
  • Ambient conditions
  • Reference equipment
  • Reference equipment specifications
  • Calibration status of the reference equipment
  • Measurement points
  • Test conditions
  • Indicated value
  • Reference value
  • Measurement difference or error
  • Measurement uncertainty
  • Method or standard used
  • Acceptance criterion, if applicable
  • Evaluation result

The more accurately this information is documented, the more useful the report will be for quality control and subsequent investigations.

How Often Should a Chamber Be Calibrated?

A fixed answer such as “annually” is not technically appropriate for all chambers.

The calibration or reevaluation interval should be determined based on the conditions of use and the equipment’s history.

Factors that may influence the determination of the interval include:

  • Frequency of chamber use
  • Operating intensity
  • Temperature range used
  • Operating conditions
  • History of failures
  • Repairs performed
  • Results of previous calibrations
  • Importance of the test
  • Customer requirements
  • Quality system requirements
  • Manufacturer’s recommendations

For example, if a chamber is used continuously and within extreme temperature ranges, the decision regarding its evaluation interval may differ from that for a chamber that is used infrequently.

Therefore, the best approach is to determine the interval based on performance history and application risk, rather than relying solely on a fixed calendar period.

When Should a Chamber Be Reexamined?

Calibration or Verification should not be limited only to a specific date on the calendar.

Certain events may create a need for reexamination, including:

  • Repair of the cooling system
  • Repair of the heating system
  • Sensor replacement
  • Controller replacement
  • Fan repair
  • Changes to the airflow system
  • Repair of the humidity system
  • Major relocation of the chamber
  • Significant changes to the control system
  • Observation of abnormal test results

For example, if the main sensor of a chamber is replaced, it is not reasonable to assume that the new system’s performance is exactly the same as before without a thorough examination.

The type of examination required depends on the nature of the repair or change.

Important Standards in the Evaluation of Environmental Test Chambers

In the field of environmental testing, the IEC 60068 series of standards is considered one of the important references.

Why Does Chamber Design Affect Calibration and Performance?

The performance of a chamber does not depend solely on sensor accuracy.

Two chambers may both use the same type of PT100 sensor, but due to differences in mechanical and thermal design, they may not have the same performance.

Factors such as:

  • Air circulation path design
  • Fan location
  • Heater location
  • Evaporator location
  • Working space volume
  • Insulation type and thickness
  • Door sealing
  • Internal design
  • Sensor location
  • Cooling capacity
  • Heating capacity
  • Control method

can all affect the uniformity and stability of the conditions inside the chamber.

For this reason, chamber accuracy is not a characteristic related solely to the sensor; it is the result of the performance of the entire system.

Why Is Air Circulation Important for Temperature Uniformity in a Chamber?

In a chamber, heat transfer does not occur only through conduction.

The air inside the chamber is circulating, and convective heat transfer takes place between the air, walls, and specimen.

If the airflow is not properly designed, some areas may:

  • Receive more heat,
  • Receive more cooling,
  • Be exposed to direct airflow,
  • Or reach the target conditions later than other locations.

Therefore, airflow path design is one of the key factors in achieving appropriate Uniformity.

From an engineering perspective, this issue is directly related to the thermal and fluid dynamics design of the chamber and cannot be resolved merely by increasing sensor accuracy.

Does Increasing the Number of Sensors Always Improve Chamber Accuracy?

No.

Increasing the number of measurement points can provide more information about the temperature or humidity distribution, but this does not, by itself, improve the chamber’s performance.

If the airflow design is not suitable, using more sensors can only reveal the existing differences more clearly.

Therefore, a distinction must be made between better performance measurement and improving the chamber’s actual performance.

The former relates to the measurement system and evaluation method; the latter relates to the design and actual performance of the device.

How Should We Select a Suitable Chamber for Accurate Testing?

When purchasing an environmental test chamber, do not examine only the temperature range.

For selecting a suitable chamber, it is also advisable to examine the following:

  • Temperature range
  • Humidity range
  • Temperature uniformity
  • Temperature stability
  • Humidity uniformity
  • Humidity stability
  • Working space volume
  • Air circulation method
  • Cooling capacity
  • Heating capacity
  • Control system
  • Sensor type
  • Sensor installation location
  • Data recording capability
  • Monitoring capability
  • Calibration and Verification conditions
  • Standards used
  • Specific test requirements

In many industrial applications, a chamber is not considered suitable merely because it “reaches the required temperature.”

What matters is its ability to create the required conditions within the working space, with appropriate stability and uniformity, and in a repeatable manner.

Common Mistakes About Chamber Calibration

  1. Considering a Calibrated Sensor Equivalent to a Calibrated Chamber

Sensor calibration only establishes the status of that particular measurement system.

  1. Paying Attention Only to the Display Value

The display may indicate the temperature at a specific point, rather than the condition of the entire working space.

  1. Treating Stability and Uniformity as the Same

Stability relates to changes over time, while Uniformity relates to spatial differences.

  1. Using a Fixed Number for Measurement Points

The number of points must be determined based on the method and the evaluation objective.

  1. Rejecting a Chamber Solely Based on the Error Value

The error must be compared with an appropriate acceptance criterion, and uncertainty must also be considered where necessary.

  1. Treating Calibration and Adjustment as the Same

Calibration is performed to determine the measurement relationship and status; Adjustment is performed to adjust or correct the system.

  1. Ignoring Loading Conditions

The performance of a chamber without a load is not necessarily the same as its performance with an actual load.

  1. Using Reference Equipment Without Checking Traceability

The measurement result must have an appropriate metrological basis.

What Does Proper Calibration Show Us?

Ultimately, the purpose of calibration and performance evaluation is not simply to produce a sheet of paper containing several numbers.

The objective is to be able to answer the following questions based on measurement data:

  • Under what conditions was the chamber examined?
  • What was the indicated value?
  • What was the reference value?
  • What was the difference?
  • What was the temperature distribution at different points?
  • How much did the temperature change over time?
  • What was the measurement uncertainty?
  • What was the acceptance criterion?
  • Do the results comply with the intended requirements?

When this information is properly documented, the measurement result becomes much more valuable for the laboratory and the quality control department.

Conclusion

Calibration of an environmental test chamber is a subject that goes beyond comparing a number on a display with a thermometer.

In a systematic evaluation, a distinction must be made between calibration of the measurement system, verification of chamber performance, and device adjustment.

Concepts such as indication error, temperature stability, temperature uniformity, humidity stability, humidity uniformity, and measurement uncertainty must also be examined separately.

Another important point is that a calibrated sensor alone does not guarantee that the entire working space of the chamber has uniform conditions. Air circulation design, the cooling and heating systems, sensor location, insulation, chamber dimensions, loading conditions, and many other factors play a role in the actual performance of the chamber.

For this reason, a test chamber with suitable performance must be designed from the outset with measurement and testing requirements in mind.

Ultimately, the purpose of calibration is not merely to know what value the chamber displays; the objective is to determine to what extent the conditions created for the test are reliable, measurable, and defensible.